·Biochemistry Analyzer
·Elisa Reader/washer
·Microwave Therapeutic
·Coagulator
·Reagent
>> 800 Tests Automatic Biochemistry Analyzer

Engine Performance

The instrument typeAutomatic random choose discrete; Emergency priority detection

testing speed800T/H1400 T/H with ISE

test methodEnd point method, two points, rate method and the immune turbidimetry/double wavelength/wavelength s

ingle wavelength method of linear and nonlinear calibration

Test project: conventional biochemical, specific protein, electrolytes, drugs, drug abuse, urine, cerebrospinal fluid, etc

analysis item at the same time: 90 photometric method, three electrode method.

Sample System

Track the style in this frameOnce every eight samples, can be on 176 samples, sample in a row

Sample plate: 27 samples, support bar code scanning (optional), independent refrigeration (4-12 ℃), to ensure quality control and calibration at any time for 24 hours a day

Sample size: 1.5 mu - 35 mu, 0.1 mul

Sample: needle with bubbles and plugging needle, liquid level detection and collision safety protection function, 

has the function of internal high pressure flushing, 1.5 mu minimum sample weight, 1.5 l - 35 mu, 0.1 mu l l progressive

Sample processing: automatic dilution, automatic measurement

Emergency samples: random insertion

Reagent System

The reagent disk90 reagentsDouble plate designDirect load70ml35ml

20mlBottom reagent bottle ,With a maximum of 4 reagent test function

Reagent refrigerated4-12 ℃ 

reagent has 24 hours continuous cold storage refrigeration function,Parr stick refrigeration

Reagent Volume15μl-300μl0.5μlstep

When testing can be deducted automatically reagent blank

probeTwo, with functions of automatic liquid level detection and collision safety protection, among them, 

the lining with functions of high pressure flushing, automatic recording reagent insufficient margin, 

or cleaning reagent needle tip function: fully automatic water inside and outside wall cleaning, t

ake all reagent pollution rate < 0.08% position fixed support bar code scanning (optional)

Reaction System

reaction disc 160new reaction cups6mmlight pathHard plastic colorimetric cup

uvPenetrating wellAcid-resistant,resistant to corrosion

The reaction liquid volume150μl -300μl

reaction temperature37The temperature accuracy±0.1℃

constant temperature systemIs a maintenance free water bath

rabbling mechanism:Two separate mixing needleLaser nano polishing

Automatic frequency conversion technology

Reaction cup cleanEight steps after the determination of samples, the instrument adopts advanced automatic cleaning, 

high pressure vacuum suction fluid, warm water wash, cleaning effect is better

Reaction cup 2 kinds of cleaning fluid

1 kind of acid, alkali, vacuum suction fluid, automatic dilution

liquid waste disposal:High and low concentration waste classification processing

With high concentration waste liquid alarm

 

Optical System

Light the lamphalogen tungsten lamplifetime2000h

After the grating spectrometer12 wavelengthes340-800nm

detectorphotodiode array

Spectroscopic methodsCluster type point source of lightAfter holographic concave grating ordinary, spectroscopic technique

 

Calibration and Quality Control

Calibration Nine calibration formulalinearA single point, two points, and more

Logit-Log4pLogit-Log5pThe sample curveexponential functionpolynomial

parabola

calibrating periodAutomatically set or set as needed

Quality control rulesWestgard Quality control rules

Cum μ Lative sum check The cumulative and rules

Twin plot Three, support for multiple levels quality control Three, 

support for multiple levels quality control

Quality control methodsQuality control, quality control in the day, days of real-time quality control

Operating System

Computer operating systemwindows 2000abovemultilanguage operating system

data processingLarge capacity storage analysis sample test resultsdisplayStore and print the response curve

Print ReportEdit and print the complete language reportOptional print content formatEditable Support network report printing

system interfacenetwork port

Work Environment

powerAC(110-220) ±22V 50HZ60HZ±1HZ

overall power consumption:≤2000VA

environment temperature;15℃-30℃

The maximum relative temperature85%

Host Volume 1860L×880W)×1200H)

 

Add: 733 Kunshan Road, Economy & Technology Development Zone of Changchun 130033, China
Tel: 86-431-84650222   E-mail:itd@sunostik.com